Gauge pressure modules have one pressure fitting and measure the pressure with respect to atmospheric pressure. Differential pressure modules have two pressure fittings and measure the difference between the applied pressure on the high fitting versus the low fitting.Each module is clearly labeled for range, over pressure and media compatibility.
Features
Model | Parameter/Range | Burst Rating6 | Hi Side Media2 | Lo Side Media2 | Reference Uncertainty4 | Total Uncertainty 1-year (15-35 °C) | Total Uncertainty 1-year1 | Total Uncertainty 6-month (15-35 °C) | Total Uncertainty 6-month1 |
---|---|---|---|---|---|---|---|---|---|
Differential | |||||||||
750P00 | 0 to 1 in H2O (0 to 2.5 mBar) | 30X | Dry Air | Dry Air | ± 0.15 % | ± 0.3 % | ± 0.35 % | ± 0.25 % | ± 0.30 % |
750P01 | 0 to 10 in H2O (0 to 25 mBar) | 3x | Dry Air | Dry Air | ± 0.1% | ± 0.2% | ± 0.3% | ± 0.15% | ± 0.25% |
750P02 | 0 to 1 psi (0 to 70 mBar) | 3x | Dry Air | Dry Air | ± 0.050% | ± 0.1% | ± 0.15% | ± 0.075% | ± 0.125% |
750P22 | 0 to 1 psi (0 to 70 mBar) | 3x | 316 SS | Dry Air | ± 0.050% | ± 0.1% | ± 0.15% | ± 0.075% | ± 0.125% |
750P03 | 0 to 5 psi (0 to 350 mBar | 3x | Dry Air | Dry Air | ± 0.02% | ± 0.04% | ± 0.05% | ± 0.035% | ± 0.04% |
750P23 | 0 to 5 psi (0 to 350 mBar) | 4x | 316 SS | Dry Air | ± 0.02% | ± 0.04% | ± 0.05% | ± 0.035% | ± 0.04% |
750P04 | 0 to 15 psi (0 to 1 bar) | 3x | Dry Air | Dry Air | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P24 | 0 to 15 psi (0 to 1 bar) | 4X | 316 SS | Dry Air | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
Gage | |||||||||
750P05 | 0 to 30 psi (0 to 2 bar) | 4x | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P06 | 0 to 100 psi (0 to 7 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P27 | 0 to 300 psi (0 to 20 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P07 | 0 to 500 psi (0 to 35 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P08 | 0 to 1000 psi (0 to 70 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P09 | 0 to 1500 psi (0 to 100 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P2000 | 0 to 2000 psi (0 to 140 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
High | |||||||||
750P29 | 0 to 3000 psi (0 to 200 bar) | 3x | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P30 | 0 to 5000 psi (0 to 340 bar) | 3x | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P31 | 0 to 10000 psi (0 to 700 bar) | 2x | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
Absolute | |||||||||
750PA3 | 0 to 5 psia (0 to 350 mBar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA4 | 0 to 15 psia (0 to 1 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA5 | 0 to 30 psia (0 to 2 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA6 | 0 to 100 psia (0 to 7 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA27 | 0 to 300 psia (0 to 20 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA7 | 0 to 500 psia (0 to 35 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA8 | 0 to 1000 psia (0 to 70 bar) | 3X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA9 | 0 to 1500 psia (0 to 100 bar) | 3X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
Vacuum | |||||||||
750PV3 | -5 psi (-350 mBar) | 4X | 316 SS | Dry Air | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PV4 | -15 psi (-1 bar) | 4X | 316 SS | Dry Air | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
Dual | |||||||||
750PD2 | -1 to 1 psi (-70 to 70 mBar) | 4X | 316 SS | Dry Air | ± 0.05% | ± 0.1% | ± 0.15% | ± 0.075% | ± 0.125% |
750PD3 | -5 to 5 psi (-350 to 350 mBar) | 4X | 316 SS | Dry Air | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PD10 | -10 to 10 psi (-700 to 700 mBar) | 4X | 316 SS | Dry Air | ± 0.025% | ± 0.05% | ± 0.07% | ± 0.04% | ± 0.06% |
750PD4 | -15 to 15 psi (-1 to 1 bar) | 4X | 316 SS | Dry Air | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD5 | -15 to 30 psi (-1 to 2 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD50 | -15 to 50 psi (-1 to 3.5 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD6 | -15 to 100 psi (-1 to 7 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD7 | -15 to 200 psi (-1 to 14 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD27 | -15 to 300 psi (-1 to 20 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
Reference | |||||||||
750R045 | 0 to 15 psi (0 to 1 bar) | 3X | Dry Air | Dry Air | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R065 | 0 to 100 psi (0 to 7 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R27 | 0 to 300 psi (0 to 20 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R07 | 0 to 500 psi (0 to 35 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R085 | 0 to 1000 psi (0 to 70 bar) | 3X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R29 | 0 to 3000 psi (0 to 200 bar) | 3X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R30 | 0 to 5000 psi (0 to 340 bar) | 3X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS | |||
750R315 | 0 to 10000 psi (0 to 700 bar) | 2X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750RD5 | -15 to 30 psi (-1 to 2 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750RD65 | -12 to 100 psi (-1 to 7 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750RD27 | -12 to 300 psi (-0.8 to 20 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
Cliquer sur une catégorie pour voir une sélection d'accessoires compatible avec le Fluke 750P Modules de pression.
Model | Parameter/Range | Burst Rating6 | Hi Side Media2 | Lo Side Media2 | Reference Uncertainty4 | Total Uncertainty 1-year (15-35 °C) | Total Uncertainty 1-year1 | Total Uncertainty 6-month (15-35 °C) | Total Uncertainty 6-month1 |
---|---|---|---|---|---|---|---|---|---|
Differential | |||||||||
750P00 | 0 to 1 in H2O (0 to 2.5 mBar) | 30X | Dry Air | Dry Air | ± 0.15 % | ± 0.3 % | ± 0.35 % | ± 0.25 % | ± 0.30 % |
750P01 | 0 to 10 in H2O (0 to 25 mBar) | 3x | Dry Air | Dry Air | ± 0.1% | ± 0.2% | ± 0.3% | ± 0.15% | ± 0.25% |
750P02 | 0 to 1 psi (0 to 70 mBar) | 3x | Dry Air | Dry Air | ± 0.050% | ± 0.1% | ± 0.15% | ± 0.075% | ± 0.125% |
750P22 | 0 to 1 psi (0 to 70 mBar) | 3x | 316 SS | Dry Air | ± 0.050% | ± 0.1% | ± 0.15% | ± 0.075% | ± 0.125% |
750P03 | 0 to 5 psi (0 to 350 mBar | 3x | Dry Air | Dry Air | ± 0.02% | ± 0.04% | ± 0.05% | ± 0.035% | ± 0.04% |
750P23 | 0 to 5 psi (0 to 350 mBar) | 4x | 316 SS | Dry Air | ± 0.02% | ± 0.04% | ± 0.05% | ± 0.035% | ± 0.04% |
750P04 | 0 to 15 psi (0 to 1 bar) | 3x | Dry Air | Dry Air | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P24 | 0 to 15 psi (0 to 1 bar) | 4X | 316 SS | Dry Air | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
Gage | |||||||||
750P05 | 0 to 30 psi (0 to 2 bar) | 4x | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P06 | 0 to 100 psi (0 to 7 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P27 | 0 to 300 psi (0 to 20 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P07 | 0 to 500 psi (0 to 35 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P08 | 0 to 1000 psi (0 to 70 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P09 | 0 to 1500 psi (0 to 100 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P2000 | 0 to 2000 psi (0 to 140 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
High | |||||||||
750P29 | 0 to 3000 psi (0 to 200 bar) | 3x | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P30 | 0 to 5000 psi (0 to 340 bar) | 3x | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P31 | 0 to 10000 psi (0 to 700 bar) | 2x | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
Absolute | |||||||||
750PA3 | 0 to 5 psia (0 to 350 mBar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA4 | 0 to 15 psia (0 to 1 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA5 | 0 to 30 psia (0 to 2 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA6 | 0 to 100 psia (0 to 7 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA27 | 0 to 300 psia (0 to 20 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA7 | 0 to 500 psia (0 to 35 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA8 | 0 to 1000 psia (0 to 70 bar) | 3X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA9 | 0 to 1500 psia (0 to 100 bar) | 3X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
Vacuum | |||||||||
750PV3 | -5 psi (-350 mBar) | 4X | 316 SS | Dry Air | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PV4 | -15 psi (-1 bar) | 4X | 316 SS | Dry Air | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
Dual | |||||||||
750PD2 | -1 to 1 psi (-70 to 70 mBar) | 4X | 316 SS | Dry Air | ± 0.05% | ± 0.1% | ± 0.15% | ± 0.075% | ± 0.125% |
750PD3 | -5 to 5 psi (-350 to 350 mBar) | 4X | 316 SS | Dry Air | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PD10 | -10 to 10 psi (-700 to 700 mBar) | 4X | 316 SS | Dry Air | ± 0.025% | ± 0.05% | ± 0.07% | ± 0.04% | ± 0.06% |
750PD4 | -15 to 15 psi (-1 to 1 bar) | 4X | 316 SS | Dry Air | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD5 | -15 to 30 psi (-1 to 2 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD50 | -15 to 50 psi (-1 to 3.5 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD6 | -15 to 100 psi (-1 to 7 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD7 | -15 to 200 psi (-1 to 14 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD27 | -15 to 300 psi (-1 to 20 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
Reference | |||||||||
750R045 | 0 to 15 psi (0 to 1 bar) | 3X | Dry Air | Dry Air | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R065 | 0 to 100 psi (0 to 7 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R27 | 0 to 300 psi (0 to 20 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R07 | 0 to 500 psi (0 to 35 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R085 | 0 to 1000 psi (0 to 70 bar) | 3X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R29 | 0 to 3000 psi (0 to 200 bar) | 3X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R30 | 0 to 5000 psi (0 to 340 bar) | 3X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS | |||
750R315 | 0 to 10000 psi (0 to 700 bar) | 2X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750RD5 | -15 to 30 psi (-1 to 2 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750RD65 | -12 to 100 psi (-1 to 7 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750RD27 | -12 to 300 psi (-0.8 to 20 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |