Ergonomic and lightweight, this multipurpose electronic air balancing instrument is ideal for taking accurate, direct air volume measurements at diffusers and grilles. This bundle includes a capture hood, a capture hood stand, and a smart tablet loaded with LogDat™ software and instruction videos.
Ergonomic and lightweight, this multipurpose electronic air balancing instrument is ideal for taking accurate, direct air volume measurements at diffusers and grilles. This bundle includes a capture hood, a capture hood stand, and a smart tablet loaded with LogDat™ software and instruction videos.
Ce produit excède les dimensions/poids de livraison standard, donc il ne se qualifie pas pour la livraison gratuite. Nous vous contacterons avec vos frais de livraisons.
Offering durable, trouble-free operation, this lightweight, ergonomically designed capture hood kit saves time and money by combining multiple measurement tools into one package. The balometer capture hood helps you create healthy and energy efficient environments while meeting local codes, guidelines and regulations for ventilation systems.
The capture hood features a detachable micromanometer which can be used with optional probes for increased flexibility in multiple measurement applications.
Features
Applications
| Velocity Range | |
| Pitot Probes | 25 to 15,500 ft/min (0.125 to 78 m/s) |
| Air Flow Probe | 25 to 5,000 ft/min (0.125 to 25 m/s) |
| Velocity Matrix | 25 to 2,500 ft/min (0.125 to 12.5 m/s) |
| Accuracy | ±3% of reading ±7 ft/min (±0.04 m/s) at velocities >50 ft/min (>0.25 m/s) |
| Resolution | 1 ft/min (0.01 m/s) |
| Pressure | |
| Differential Pressure | ±15 inH2O (±3735 Pa) 150 inH2O (37.5 kPa) Maximum safe operating pressure |
| Absolute Pressure | 15 to 40 inHg (356 to 1016 mmHg) |
| Accuracy | ±2% of reading ±0.0001 inH2O (±0.025 Pa) static and differential ±2% of reading absolute |
| Resolution | 0.00001 inH2O (0.001 Pa) static and differential 0.01 inHg (1 mmHg) absolute |
| General Specifications | |
| Instrument Temperature Range | Operating: 40 to 140°F (4.4 to 60°C) Storage: -4 to 160°F (-20 to 71°C) |
| Weight | 7.4 lbs (3.4 kg) |
Cliquer sur une catégorie pour voir une sélection d'accessoires compatible avec le TSI EBT731-STA Balometer Capture Hood with stand and tablet.
| Velocity Range | |
| Pitot Probes | 25 to 15,500 ft/min (0.125 to 78 m/s) |
| Air Flow Probe | 25 to 5,000 ft/min (0.125 to 25 m/s) |
| Velocity Matrix | 25 to 2,500 ft/min (0.125 to 12.5 m/s) |
| Accuracy | ±3% of reading ±7 ft/min (±0.04 m/s) at velocities >50 ft/min (>0.25 m/s) |
| Resolution | 1 ft/min (0.01 m/s) |
| Pressure | |
| Differential Pressure | ±15 inH2O (±3735 Pa) 150 inH2O (37.5 kPa) Maximum safe operating pressure |
| Absolute Pressure | 15 to 40 inHg (356 to 1016 mmHg) |
| Accuracy | ±2% of reading ±0.0001 inH2O (±0.025 Pa) static and differential ±2% of reading absolute |
| Resolution | 0.00001 inH2O (0.001 Pa) static and differential 0.01 inHg (1 mmHg) absolute |
| General Specifications | |
| Instrument Temperature Range | Operating: 40 to 140°F (4.4 to 60°C) Storage: -4 to 160°F (-20 to 71°C) |
| Weight | 7.4 lbs (3.4 kg) |
Cliquer sur une catégorie pour voir une sélection d'accessoires compatible avec le TSI EBT731-STA Balometer Capture Hood with stand and tablet.