Waterproof and dustproof with IP67 rated housing that floats! Designed for portability and durability the waterproof T-100 turbidity meter is ideal for monitoring turbidity in chemical, food and industrial applications.
Waterproof and dustproof with IP67 rated housing that floats! Designed for portability and durability the waterproof T-100 turbidity meter is ideal for monitoring turbidity in chemical, food and industrial applications.
The microprocessor-based T-100 turbidity meter uses an infrared LED light source and delivers unprecedented repeatability and accuracy while offering resolution as low as 0.01 NTU. This lightweight meter is a valuable analytical tool for field-testing and quality control.
Features
| Range Selection | 0.01 to 19.99 NTU, 20.0 to 99.9 NTU, 100 to 1000 NTU |
| Resolution | 0.01 NTU, 0.1 NTU, 1 NTU |
| Accuracy | ±2%: 0 to 500 NTU; ±3%: 201 to 1000 NTU |
| Repeatability | ±0.01 NTU or ±1%, whichever is greater |
| Response Time | < 6 seconds for full step change |
| Calibration Standards | 0.02, 20.0, 100, 800 NTU |
| Light Source | Infrared-emitting diode, 850 nm |
| Light Source Life | >1,000,000 measurements |
| Stray Light | < 0.02 NTU |
| Detector | Silicon photovoltaic |
| Display | 4-digit, 14-segment LCD |
| Sample Vials | Borosilicate glass w/ screw caps, fill line & indexing mark; 2"H x 1" dia. (51 x 25mm) |
| Sample Volume | 10 ml (0.33 oz) |
| Power Supply | 4 x "AAA" alkaline batteries |
| Dimensions/Weight | 6.25 x 2.75 x 2" (155 x 68 x 46mm)/7oz (200g) |
Cliquer sur une catégorie pour voir une sélection d'accessoires compatible avec le OAKTON T-100 WD-35635-05 Turbidity Meter.
| Range Selection | 0.01 to 19.99 NTU, 20.0 to 99.9 NTU, 100 to 1000 NTU |
| Resolution | 0.01 NTU, 0.1 NTU, 1 NTU |
| Accuracy | ±2%: 0 to 500 NTU; ±3%: 201 to 1000 NTU |
| Repeatability | ±0.01 NTU or ±1%, whichever is greater |
| Response Time | < 6 seconds for full step change |
| Calibration Standards | 0.02, 20.0, 100, 800 NTU |
| Light Source | Infrared-emitting diode, 850 nm |
| Light Source Life | >1,000,000 measurements |
| Stray Light | < 0.02 NTU |
| Detector | Silicon photovoltaic |
| Display | 4-digit, 14-segment LCD |
| Sample Vials | Borosilicate glass w/ screw caps, fill line & indexing mark; 2"H x 1" dia. (51 x 25mm) |
| Sample Volume | 10 ml (0.33 oz) |
| Power Supply | 4 x "AAA" alkaline batteries |