Featuring reference-grade platinum sensors, this secondary PRT with spade lugs also features 4-wire connections with less noisy measurements than its 2-wire counterparts.
Featuring reference-grade platinum sensors, this secondary PRT with spade lugs also features 4-wire connections with less noisy measurements than its 2-wire counterparts.
Features
| Temperature Range | -328 to 1,238°F (-200 to 670°C) | 
| Nominal Resistance at 0.01°C | 100Ω ±0.5Ω | 
| Temperature Coefficient | 0.0039250Ω/Ω/°C | 
| Short-Term Repeatability | ±0.01 at 0.010°C; ±0.02°C at max temp | 
| Drift | ±0.01 at 0.010°C; ±0.02°C at max temp | 
| Hysteresis | ±0.01°C maximum | 
| Sensor Length | 1.2in ±0.2in (30mm ±5mm) | 
| Sensor Location | 0.1in ±0.1in (3mm ±1mm from tip) | 
| Minimum Insulation Resistance | 500MΩ at 23°C, 10MΩ at 670°C | 
| Transition Junction Dimensions | 2.8 x 0.49in (71 x 12.5mm) | 
| Minimum Immersion Length (<5 mK Error) | 3.9in (100mm) | 
| Maximum Immersion Length | 12in (305mm) | 
| Response Time | 12 seconds typical | 
| Self Heating (In 0°C Bath) | 50mW/°C | 
| Lead-Wire Cable Type | Teflon, 24AWG | 
| Lead-Wire Length | 6ft (1.8m) | 
| Lead-Wire Temperature Range | -58 to 482°F (-50 to 250°C) | 
| Calibration | Calibration not included; NVLAP-accredited calibration optional, lab code 200348-0; please see calibration uncertainty table and its explanation of changeable uncertainties | 
Cliquer sur une catégorie pour voir une sélection d'accessoires compatible avec le Fluke 5609-12-S Secondary PRT with Spade Lugs, -200 to 670°C, Calibration Options.
| Temperature Range | -328 to 1,238°F (-200 to 670°C) | 
| Nominal Resistance at 0.01°C | 100Ω ±0.5Ω | 
| Temperature Coefficient | 0.0039250Ω/Ω/°C | 
| Short-Term Repeatability | ±0.01 at 0.010°C; ±0.02°C at max temp | 
| Drift | ±0.01 at 0.010°C; ±0.02°C at max temp | 
| Hysteresis | ±0.01°C maximum | 
| Sensor Length | 1.2in ±0.2in (30mm ±5mm) | 
| Sensor Location | 0.1in ±0.1in (3mm ±1mm from tip) | 
| Minimum Insulation Resistance | 500MΩ at 23°C, 10MΩ at 670°C | 
| Transition Junction Dimensions | 2.8 x 0.49in (71 x 12.5mm) | 
| Minimum Immersion Length (<5 mK Error) | 3.9in (100mm) | 
| Maximum Immersion Length | 12in (305mm) | 
| Response Time | 12 seconds typical | 
| Self Heating (In 0°C Bath) | 50mW/°C | 
| Lead-Wire Cable Type | Teflon, 24AWG | 
| Lead-Wire Length | 6ft (1.8m) | 
| Lead-Wire Temperature Range | -58 to 482°F (-50 to 250°C) | 
| Calibration | Calibration not included; NVLAP-accredited calibration optional, lab code 200348-0; please see calibration uncertainty table and its explanation of changeable uncertainties | 
Cliquer sur une catégorie pour voir une sélection d'accessoires compatible avec le Fluke 5609-12-S Secondary PRT with Spade Lugs, -200 to 670°C, Calibration Options.