Generates pressure to an accuracy of 0.1% of reading. These deadweight testers quality as primary standards for pressure calibration.
Generates pressure to an accuracy of 0.1% of reading. These deadweight testers quality as primary standards for pressure calibration.
Ce produit excède les dimensions/poids de livraison standard, donc il ne se qualifie pas pour la livraison gratuite. Nous vous contacterons avec vos frais de livraisons.
Comes equipped with a built-in shuttle valve which provides the means to control the rate of pressure increase, while the precision adjustment is accomplished with an integrated micro vernier displacement valve. This deadweight tester also provides a precise means for generating pressure with high accuracy that can be used as a primary calibration standard.
Features
Applications
Accuracy | Within 0.1% of produced pressure (adjusted for standard gravity conditions; compensation for local gravity effects may be required) |
Operating Pressure | 15 psi to 10,000 psi maximum |
Reservoir Volume | Approximately 1.5 pints (0.7 l) |
Pump Body Material | Aluminum alloy |
Piston and Cylinder Material | Piston: 440C stainless steel Cylinder: 416 stainless steel |
Operating Fluid | SAE 20 automotive or machine oil |
Weight Material | Non-magnetic die cast zinc alloy, ASTM AC41A |
O-Ring Seals | Buna-N |
Certification | Standard: NIST traceability document provided with each unit |
Accuracy | Within 0.1% of produced pressure (adjusted for standard gravity conditions; compensation for local gravity effects may be required) |
Operating Pressure | 15 psi to 10,000 psi maximum |
Reservoir Volume | Approximately 1.5 pints (0.7 l) |
Pump Body Material | Aluminum alloy |
Piston and Cylinder Material | Piston: 440C stainless steel Cylinder: 416 stainless steel |
Operating Fluid | SAE 20 automotive or machine oil |
Weight Material | Non-magnetic die cast zinc alloy, ASTM AC41A |
O-Ring Seals | Buna-N |
Certification | Standard: NIST traceability document provided with each unit |