Measuring DC voltage/current and source DC mA current with an accuracy of 0.02%, this process calibrator can simultaneously provide 24V loop power to transmitter while conducting mA current measurement. In addition, it features an adjustable 0-20mA/4-20mA span with % readout, automatic sawtooth or triangular output current step function, auto power off, and a built-in flashlight.
Measuring DC voltage/current and source DC mA current with an accuracy of 0.02%, this process calibrator can simultaneously provide 24V loop power to transmitter while conducting mA current measurement. In addition, it features an adjustable 0-20mA/4-20mA span with % readout, automatic sawtooth or triangular output current step function, auto power off, and a built-in flashlight.
The Triplett PCAL300 Loop Process Calibrator for benchtop and field industrial process device testing and calibration features an adjustable 0-20mA/4-20mA span with % readout, automatic sawtooth or triangular output current step function, auto power off, and a built-in flashlight.
Features
Applications
| Measurement | |
| DC Voltage | 30 V (0.001 V) Accuracy: ±(0.02% + 2 mV) |
| DC Current | 30 mA (0.001 mA) Accuracy: ±(0.02% + 4μA) |
| Source | |
| DC Current | 30 mA (0.001 mA) Accuracy: ±(0.05% + 4μA) |
| Loop Power | 24 V Accuracy: ±(10%) |
| General Specifications | |
| Power | 3 x AA batteries |
| Dimensions | 7 x 3.5 x 2" (178 x 89 x 51 mm) |
| Weight | 1 lbs (454 g) |
| Measurement | |
| DC Voltage | 30 V (0.001 V) Accuracy: ±(0.02% + 2 mV) |
| DC Current | 30 mA (0.001 mA) Accuracy: ±(0.02% + 4μA) |
| Source | |
| DC Current | 30 mA (0.001 mA) Accuracy: ±(0.05% + 4μA) |
| Loop Power | 24 V Accuracy: ±(10%) |
| General Specifications | |
| Power | 3 x AA batteries |
| Dimensions | 7 x 3.5 x 2" (178 x 89 x 51 mm) |
| Weight | 1 lbs (454 g) |