Built for direct-to-sensor measurement, this USB data acquisition system connects sensors directly to a computer over a USB connection. It provides 16 universal analog inputs and 4 digital I/O lines. It samples at up to 160,000 S/s, provides high-accuracy resolution, and is powered over the USB bus. Best for multi-channel sensor monitoring and logging.

Built for direct-to-sensor measurement, this USB data acquisition system connects sensors directly to a computer over a USB connection. It provides 16 universal analog inputs and 4 digital I/O lines. It samples at up to 160,000 S/s, provides high-accuracy resolution, and is powered over the USB bus. Best for multi-channel sensor monitoring and logging.

Your Price $1332.00 CAD
Availability 2 to 3 Weeks
Quantity

OMEGA INET-600 Offers

The iNET-600 Series A/D module provides 16SE/8DI voltage input channels (Ch#1...#16), each of which are independently software programmable with Windows software that support the direct connection to many common sensor types.

Alternatively, you can attach iNET-600 Series to the following optional wiring boxes: iNET-510, iNET-511, iNET-512. If you are working with thermocouples, an iNET-510 wiring box is required due to its internal cold junction compensation.

iNET-600 Series is a stand-alone USB data acquisition system. No additional components, such as external power supply, are required.

Each channel provides the following software programmable parameters: A/D Signal-Averaging-Per-Point (0 to 100 mSec), Sample-Rate (samples-per-second-per-channel), Digital IIR Filter (Low Pass, High Pass, Band Pass, or Band Stop), Voltage Measurement Range (±20 mV to ±10V), Sensor Type, and Single-Ended or Differential Wiring.

Excitation power (+3.3V ±0.2V, <80 mA, 28 mA per sensor max) is provided for sensors, along with other End User Power voltages. This 3.3V, which is referenced to instruNet Ground, is automatically read back by A/D when calculating sensor values.

The 4 mA sink/source digital I/O port consists of 4 individual TTL-compatible lines (Ch#25...#28), each of which can be configured as: input or output bit. When configured as an input, a channel can be used to sense a digital high (2 to 5.5V) or digital low (0V to 0.8V). When configured as an output, a channel can be set high (e.g. >2V) or low (e.g. <0.8V).

These I/O pins are short-circuit protected against high voltages up to 6.0V and down to -6.0V.

Waveforms are recorded in one long continuous stream (strip chart mode); or they are digitized in short segments where each segment begins after a specified trigger event (oscilloscope mode). Any one of the instruNet analog or digital input channels can be used as a trigger.

Data is either placed into RAM memory, a file on disk, or is deleted after being plotted. One uses the Setup dialog box to set the sample rate (i.e. points-digitized-per-second-per-channel) and number of points digitized; and uses the Network page to select which channels are digitized.

instruNet World Software is a FREE application program that manages, monitors and operates instruNet data acquisition and control hardware. Additionally, it digitizes long continuous waveforms, spools them to disk, views incoming waveforms in real-time and then allows post acquisition viewing, like an oscilloscope or strip chart recorder.

instruNet is compatible with Excel, LabVIEW(R), DASYLab(R), MATLAB, Origin, C & Visual BASIC software. Application software is used to configure all I/O channels, store your settings, view digitized data in real-time, analyze, stream data to disk, and scroll through your waveform post-acquisition.

Features

  • 16SE/8DI voltage input channels with extremely accurate 24-bit A/D converter that supports ±20 mV to ±10V ranges
  • 4 digital I/O (4 mA sink/source, 0 to 3.3V)
  • Controls process via Windows(R) software using analog inputs and digital I/O
  • iNET-600 connects I/O signal ground to computer ground
  • Receives power from the USB bus and provides excitation power (+3.3V, 80 mA) to sensors
  • Easily set-up sensors using the included instruNet World software
  • Connect a computer directly to sensors: voltage, thermocouple, thermistor, RTD, load cell, strain gage, potentiometer, current, resistance

Applications

  • Direct-to-sensor temperature monitoring with thermocouples, RTDs and thermistors
  • Strain gage and load cell measurement for structural and weighing test setups
  • Real-time waveform recording and plotting in strip chart or oscilloscope mode
  • Streaming digitized sensor data directly into Excel or LabVIEW for analysis
  • Process control loops using analog inputs and digital I/O via Windows software

OMEGA INET-600 Specifications

Number of analog channels 16 single-ended (SE)/8 differential (DI)
A/D converter 24-bit when averaging each point for >=1 msec; and 16 bits when not averaging
Voltage ranges ±20 mV, ±40 mV, ±80 mV, ±150 mV, ±300 mV, ±600 mV, ±1.2 V, ±2.5 V, ±5 V, ±10 V
Protected voltage ±30 Vdc with power on or off
RFI filter 13 KHz RFI filter on <= ±150 mV range
Digital filter Low Pass, High Pass, Band Pass, or Band Stop
Digital filter type Elliptic, Chebyshev B, Chebyshev S, or Butterworth
Common mode voltage ±10 Vdc
Common mode rejection ratio >=110 dB (0 to 60 Hz)
Input impedance 100 MΩ
Thermocouple J/K/T/E/R/S/B/N/C/G/D
Thermistor 2252 Ω (OMEGA 44000 Series), custom
RTD 100 Ω, 500 Ω, 1 kΩ platinum
Load cell 10 kg, 25 kg, 100 kg, 250 kg, 1000 kg, 5000 kg, custom
Strain 120 Ω or 350 Ω. 1/4 bridge, half bridge or full bridge; internal i60x voltage reference supports a maximum of three 120 Ω bridges; and eight 350 Ω bridges
Potentiometer 1 to 50 kΩ
Current ±80 uA, ±120 uA, ±500 uA, ±600 uA, ±800 uA, ±1.2 mA, ±2.5 mA, ±12 mA, ±24 mA, 0 to 24 mA, custom
Resistance 0 to 33 Ω, 0 to 100 Ω, 0 to 330 Ω, 0 to 1 kΩ, 0 to 3300 Ω, 0 to 10 kΩ, custom
LVDT/Flow/Pressure Current, voltage, mV/V
Number of digital channels 4 digital I/O (4 mA sink/source, 0 to 3.3 V)
"0" input voltage 0 to 0.8 Vdc
"1" input voltage 2 to 5.5 Vdc
"1" input current Amps = Vin / 50 kΩ ("1" must drive internal 50 kΩ pull-down resistor)
"0" output voltage <0.7 Vdc @ <2 mA; <0.9 Vdc @ 4 mA
"1" output voltage 2.2 to 3.3 Vdc @ 4 mA
Pull-down resistor 50 kΩ ±1%
Operating environment 34 to 158°F (1 to 70°C), <90% RH, no condensation
Storage temperature -4 to 158°F (-20 to 70°C)
Power consumed from USB cable +5 Vdc 255 mA typical (433 mA max)
Voltage reference at Hd44 connector 3.30 Vdc (80 mA)
Unregulated power available to user at Hd44 connector 5 Vdc (8 mA), ±15 Vdc (4 mA)
Material Plastic enclosure
Mounting Place on desktop, attach to wall, or place inside another enclosure (OEM)
Description Standalone direct-to-sensor USB data acquisition system, 16SE/8DI voltage inputs, 4 digital I/O
Dimensions 3.784 x 0.924 x 2.286" (58 H x 96 W x 24 mm D)
Weight 1 lb (454 g)

What's included with the OMEGA INET-600

  • Direct-to-Sensor USB Data Acquisition System
  • USB Interface Cable
  • instruNet World Software on CD
  • Mating Hd44 Female Connector
  • Hd44 Cover
  • Hardware Assembly Kit

Ask a question about Omega INET-600 USB Data Acquisition System, 16 Single-Ended or 8 Differential Universal Inputs, 160,000 S/s

Customer Reviews for the OMEGA INET-600

OMEGA INET-600 Offers

The iNET-600 Series A/D module provides 16SE/8DI voltage input channels (Ch#1...#16), each of which are independently software programmable with Windows software that support the direct connection to many common sensor types.

Alternatively, you can attach iNET-600 Series to the following optional wiring boxes: iNET-510, iNET-511, iNET-512. If you are working with thermocouples, an iNET-510 wiring box is required due to its internal cold junction compensation.

iNET-600 Series is a stand-alone USB data acquisition system. No additional components, such as external power supply, are required.

Each channel provides the following software programmable parameters: A/D Signal-Averaging-Per-Point (0 to 100 mSec), Sample-Rate (samples-per-second-per-channel), Digital IIR Filter (Low Pass, High Pass, Band Pass, or Band Stop), Voltage Measurement Range (±20 mV to ±10V), Sensor Type, and Single-Ended or Differential Wiring.

Excitation power (+3.3V ±0.2V, <80 mA, 28 mA per sensor max) is provided for sensors, along with other End User Power voltages. This 3.3V, which is referenced to instruNet Ground, is automatically read back by A/D when calculating sensor values.

The 4 mA sink/source digital I/O port consists of 4 individual TTL-compatible lines (Ch#25...#28), each of which can be configured as: input or output bit. When configured as an input, a channel can be used to sense a digital high (2 to 5.5V) or digital low (0V to 0.8V). When configured as an output, a channel can be set high (e.g. >2V) or low (e.g. <0.8V).

These I/O pins are short-circuit protected against high voltages up to 6.0V and down to -6.0V.

Waveforms are recorded in one long continuous stream (strip chart mode); or they are digitized in short segments where each segment begins after a specified trigger event (oscilloscope mode). Any one of the instruNet analog or digital input channels can be used as a trigger.

Data is either placed into RAM memory, a file on disk, or is deleted after being plotted. One uses the Setup dialog box to set the sample rate (i.e. points-digitized-per-second-per-channel) and number of points digitized; and uses the Network page to select which channels are digitized.

instruNet World Software is a FREE application program that manages, monitors and operates instruNet data acquisition and control hardware. Additionally, it digitizes long continuous waveforms, spools them to disk, views incoming waveforms in real-time and then allows post acquisition viewing, like an oscilloscope or strip chart recorder.

instruNet is compatible with Excel, LabVIEW(R), DASYLab(R), MATLAB, Origin, C & Visual BASIC software. Application software is used to configure all I/O channels, store your settings, view digitized data in real-time, analyze, stream data to disk, and scroll through your waveform post-acquisition.

Features

  • 16SE/8DI voltage input channels with extremely accurate 24-bit A/D converter that supports ±20 mV to ±10V ranges
  • 4 digital I/O (4 mA sink/source, 0 to 3.3V)
  • Controls process via Windows(R) software using analog inputs and digital I/O
  • iNET-600 connects I/O signal ground to computer ground
  • Receives power from the USB bus and provides excitation power (+3.3V, 80 mA) to sensors
  • Easily set-up sensors using the included instruNet World software
  • Connect a computer directly to sensors: voltage, thermocouple, thermistor, RTD, load cell, strain gage, potentiometer, current, resistance

Applications

  • Direct-to-sensor temperature monitoring with thermocouples, RTDs and thermistors
  • Strain gage and load cell measurement for structural and weighing test setups
  • Real-time waveform recording and plotting in strip chart or oscilloscope mode
  • Streaming digitized sensor data directly into Excel or LabVIEW for analysis
  • Process control loops using analog inputs and digital I/O via Windows software

OMEGA INET-600 Specifications

Number of analog channels 16 single-ended (SE)/8 differential (DI)
A/D converter 24-bit when averaging each point for >=1 msec; and 16 bits when not averaging
Voltage ranges ±20 mV, ±40 mV, ±80 mV, ±150 mV, ±300 mV, ±600 mV, ±1.2 V, ±2.5 V, ±5 V, ±10 V
Protected voltage ±30 Vdc with power on or off
RFI filter 13 KHz RFI filter on <= ±150 mV range
Digital filter Low Pass, High Pass, Band Pass, or Band Stop
Digital filter type Elliptic, Chebyshev B, Chebyshev S, or Butterworth
Common mode voltage ±10 Vdc
Common mode rejection ratio >=110 dB (0 to 60 Hz)
Input impedance 100 MΩ
Thermocouple J/K/T/E/R/S/B/N/C/G/D
Thermistor 2252 Ω (OMEGA 44000 Series), custom
RTD 100 Ω, 500 Ω, 1 kΩ platinum
Load cell 10 kg, 25 kg, 100 kg, 250 kg, 1000 kg, 5000 kg, custom
Strain 120 Ω or 350 Ω. 1/4 bridge, half bridge or full bridge; internal i60x voltage reference supports a maximum of three 120 Ω bridges; and eight 350 Ω bridges
Potentiometer 1 to 50 kΩ
Current ±80 uA, ±120 uA, ±500 uA, ±600 uA, ±800 uA, ±1.2 mA, ±2.5 mA, ±12 mA, ±24 mA, 0 to 24 mA, custom
Resistance 0 to 33 Ω, 0 to 100 Ω, 0 to 330 Ω, 0 to 1 kΩ, 0 to 3300 Ω, 0 to 10 kΩ, custom
LVDT/Flow/Pressure Current, voltage, mV/V
Number of digital channels 4 digital I/O (4 mA sink/source, 0 to 3.3 V)
"0" input voltage 0 to 0.8 Vdc
"1" input voltage 2 to 5.5 Vdc
"1" input current Amps = Vin / 50 kΩ ("1" must drive internal 50 kΩ pull-down resistor)
"0" output voltage <0.7 Vdc @ <2 mA; <0.9 Vdc @ 4 mA
"1" output voltage 2.2 to 3.3 Vdc @ 4 mA
Pull-down resistor 50 kΩ ±1%
Operating environment 34 to 158°F (1 to 70°C), <90% RH, no condensation
Storage temperature -4 to 158°F (-20 to 70°C)
Power consumed from USB cable +5 Vdc 255 mA typical (433 mA max)
Voltage reference at Hd44 connector 3.30 Vdc (80 mA)
Unregulated power available to user at Hd44 connector 5 Vdc (8 mA), ±15 Vdc (4 mA)
Material Plastic enclosure
Mounting Place on desktop, attach to wall, or place inside another enclosure (OEM)
Description Standalone direct-to-sensor USB data acquisition system, 16SE/8DI voltage inputs, 4 digital I/O
Dimensions 3.784 x 0.924 x 2.286" (58 H x 96 W x 24 mm D)
Weight 1 lb (454 g)

What's included with the OMEGA INET-600

  • Direct-to-Sensor USB Data Acquisition System
  • USB Interface Cable
  • instruNet World Software on CD
  • Mating Hd44 Female Connector
  • Hd44 Cover
  • Hardware Assembly Kit

Ask a question about Omega INET-600 USB Data Acquisition System, 16 Single-Ended or 8 Differential Universal Inputs, 160,000 S/s