The AVS140-6 is a complete system designed for autoclave validations. This model includes 6 x HiTemp140-1 data loggers, IFC406 interface cable, license for MadgeTech’s FDA 21 CFR Part 11 Secure Software, and IQ/OQ/PQ validation protocols.
The AVS140-6 is a complete system designed for autoclave validations. This model includes 6 x HiTemp140-1 data loggers, IFC406 interface cable, license for MadgeTech’s FDA 21 CFR Part 11 Secure Software, and IQ/OQ/PQ validation protocols.
HiTemp140
The HiTemp140 data logger is MadgeTech’s solution for high precision, rugged temperature monitoring. It can withstand temperatures from -40 to 284°F (-40 to 140°C) and has an accuracy of ±0.18°F (±0.1°C).
The HiTemp140 can store up to 32,256 readings, and features a rigid external probe capable of measuring extended temperatures, up to 500°F (260°C). Custom probe lengths up to 7" (17 cm) are available. Also available, is the HiTemp140-PT that features a 24" (61 cm) flexible stainless steel probe for measuring extended temperatures up to 662°F (350°C). The probe is durable and can be spiraled, bent or angled in any direction, making it easy to log temperatures within bottles, vials or other hard to reach places.
PR140
The PR140 is a data logger designed to validate if appropriate pressure levels have been achieved during the steam sterilization cycle. It is built with a precision stainless steel pressure gauge, the data logger has an accuracy of ±0.03 Bar (±0.435 PSI), which can be achieved over a wide temperature range, from 68 to 284°F (20 to 140°C). The PR140 is available in either a Flush Top or NPT Pressure Port Top design, with an optional female luer fitting accessory.
IFC406
The IFC406 Multiplexer Data Logger Interface allows for multiple devices to be connected into one interface. Each IFC406 allows for 6 data loggers to be connected.
MadgeTech 4 Secure Software
The MadgeTech 4 Secure Software aids customers in compliance with 21 CFR Part 11 requirements. The software ensures standards in which electronic files are considered equivalent to paper records, saving time and effort. MadgeTech 4 Secure Software contains criteria such as electronic signatures, access codes, secure data files, and an audit trail which meet the requirements of 21 CFR Part 11 and help provide data integrity
MadgeTech provides on-site IQ/OQ/PQ services to help the user validate the data logging system. Our trained staff are extremely knowledgeable with the standard IQ/OQ/PQ protocols of the MadgeTech data logging system. This service is a cost-effective means to save time and ensure proper implementation of IQ/OQ/PQ of the MadgeTech data logging system with a minimal disruption to staff and operations. Get more information on our IQ/OQ/PQ On-Site Service.
Applications
HiTemp140 | |
Temperature Sensor | 100 Ω Platinum RTD |
Probe Measurement Range | HiTemp140: -328 to 500°F (-200 to 260°C) HiTemp140-PT: -328 to 662°F (-200 to 350°C) |
Temperature Resolution | 0.02°F (0.01°C) |
Calibrated Accuracy | ±0.18°F (±0.1°C) 68 to 284°F (20 to 140°C) ±0.54°F (±0.3°C) -4 to 67.98°F (-20 to 19.99°C) ±0.72°F (±0.4°C) -40 to -4.02°F (-40 to -20.01°C) |
Reading Rate | 1 second up to once every 24 hours |
Memory | 32,256 readings |
Battery Life | 2 years typical (1 minute reading rate at 77°F (25°C)) |
Time Accuracy | ±1 minute/month at 68 to 86°F (20 to 30°C) Stand alone mode |
Operating Environment | -40 to 284°F (-40 to 140°C) 0 to 100% E12RH |
Submersible | Yes (IP68) |
Material | 316 Stainless Steel |
Approvals | CE |
Dimensions | 1.6 x 0.970" diameter (40 x 24.6 mm diameter) |
Weight | 4.2 oz (120 g) |
PR140 | |
Pressure Sensor | Semiconductor (Strain Gauge) |
Pressure Range | 0 to 5 Bar (0 to 72.5 PSIA) |
Pressure Resolution | 0.1 mBar (0.0015 PSIA) |
Calibrated Accuracy | ±0.03 Bar (±0.435 PSI) 68 to 284°F (20 to 140°C) |
Response Time | 0.1 ms (10 to 90% FSR) |
Reading Rate | 1 second up to once every 24 hours |
Memory | 32,256 readings |
Battery Life | 2 years typical, 1 minute reading rate at 77°F (25°C) |
Time Accuracy | ±1 minutes/month at 68°F (20°C) |
Operating Environment | -4 to 284°F (-20 to +140°C) 0 to 100% RH Note: May be used above 140°F (60°C) for up to 8 hours per 24-hour period |
Submersible | Yes (IP68) |
Material | Stainless Steel |
Approvals | CE |
Dimensions | 2.3 x 1.0" diameter (58.2 x 25.4 mm diameter) |
Weight | 3 oz (85 g) |
IFC406 | |
Connection Type | USB (to PC) |
Material | 6061 Aluminum (PTFE impregnated hard anodized coating) ABS Plastic |
Maximum Input Voltage | 6 Volts (V) |
Approval | CE |
Operating Environment | 50 to 95°F (10 to 35°C) 0 to 95% RH non-condensing |
Battery | This device does not use a battery |
Dimensions | 9.5 x 1.95 x 1.75" (241.3 x 49.5 x 44.5 mm) |
Weight | 1.65 lbs (750 g) |
HiTemp140 | |
Temperature Sensor | 100 Ω Platinum RTD |
Probe Measurement Range | HiTemp140: -328 to 500°F (-200 to 260°C) HiTemp140-PT: -328 to 662°F (-200 to 350°C) |
Temperature Resolution | 0.02°F (0.01°C) |
Calibrated Accuracy | ±0.18°F (±0.1°C) 68 to 284°F (20 to 140°C) ±0.54°F (±0.3°C) -4 to 67.98°F (-20 to 19.99°C) ±0.72°F (±0.4°C) -40 to -4.02°F (-40 to -20.01°C) |
Reading Rate | 1 second up to once every 24 hours |
Memory | 32,256 readings |
Battery Life | 2 years typical (1 minute reading rate at 77°F (25°C)) |
Time Accuracy | ±1 minute/month at 68 to 86°F (20 to 30°C) Stand alone mode |
Operating Environment | -40 to 284°F (-40 to 140°C) 0 to 100% E12RH |
Submersible | Yes (IP68) |
Material | 316 Stainless Steel |
Approvals | CE |
Dimensions | 1.6 x 0.970" diameter (40 x 24.6 mm diameter) |
Weight | 4.2 oz (120 g) |
PR140 | |
Pressure Sensor | Semiconductor (Strain Gauge) |
Pressure Range | 0 to 5 Bar (0 to 72.5 PSIA) |
Pressure Resolution | 0.1 mBar (0.0015 PSIA) |
Calibrated Accuracy | ±0.03 Bar (±0.435 PSI) 68 to 284°F (20 to 140°C) |
Response Time | 0.1 ms (10 to 90% FSR) |
Reading Rate | 1 second up to once every 24 hours |
Memory | 32,256 readings |
Battery Life | 2 years typical, 1 minute reading rate at 77°F (25°C) |
Time Accuracy | ±1 minutes/month at 68°F (20°C) |
Operating Environment | -4 to 284°F (-20 to +140°C) 0 to 100% RH Note: May be used above 140°F (60°C) for up to 8 hours per 24-hour period |
Submersible | Yes (IP68) |
Material | Stainless Steel |
Approvals | CE |
Dimensions | 2.3 x 1.0" diameter (58.2 x 25.4 mm diameter) |
Weight | 3 oz (85 g) |
IFC406 | |
Connection Type | USB (to PC) |
Material | 6061 Aluminum (PTFE impregnated hard anodized coating) ABS Plastic |
Maximum Input Voltage | 6 Volts (V) |
Approval | CE |
Operating Environment | 50 to 95°F (10 to 35°C) 0 to 95% RH non-condensing |
Battery | This device does not use a battery |
Dimensions | 9.5 x 1.95 x 1.75" (241.3 x 49.5 x 44.5 mm) |
Weight | 1.65 lbs (750 g) |