A user can accurately determine the concentration of chlorine over two ranges, low range and a high range mode. Using the HI93701-01 reagents for low range and the HI95771-01 for high range.
A user can accurately determine the concentration of chlorine over two ranges, low range and a high range mode. Using the HI93701-01 reagents for low range and the HI95771-01 for high range.
Discontinued!
This product has been discontinued and is no longer available.
This is the direct replacement:
HI 96771 is a photometer developed to check chlorine dosing in disinfection procesesses with high concentrations ofchlorine. Thanks to the extended range from 0 to 500 ppm, it is ideal for the food industry, such as in fruit and vegetable washing.
With the powerful CAL CHECK validation function, users are able to validate performance of the instruments at any time. The validation procedure is extremely user friendly. Just use the exclusive HANNA ready-made, NIST traceable standards to verify the performance of the instruments and recalibrate if necessary.
The instruments are water resistant and the lamp and filter units are protected from dust or dirt by a transparent cup.
These meters use an exclusive positive-locking system to ensure that the cuvet is in the same position every time it is placed into the measurement cell. It is designed to fit a cuvet with a larger neck making it easier to add both sample and reagents. The cuvet is made from special optical glass to obtain best results.
The reagents are in powder and liquid form and are supplied in packets and bottles respectively. The amount of reagent is precisely dosed to ensure the maximum repeatability.
Features
| Range | Free Chlorine (P1): 0.00 to 5.00 mg/L (ppm); Ultra High Range Chlorine (P2): 0 to 500 mg/L (ppm) |
| Resolution | Free Chlorine (P1): 0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/L; Ultra High Range Chlorine (P2): 1 mg/L from 0 to 200 mg/L; 10 mg/L above 200 mg/L |
| Accuracy @ 25°C (77°F) | Free Chlorine (P1): ±0.03 mg/L ±3% of reading; Ultra High Range Chlorine (P2): ±3 mg/L ±3% of reading |
| Method | adaptation of the Standard Methods for the Examination of Water and Wastewater, 4500-Cl method |
| Auto-off | after ten minutes of non-use in measurement mode; after one hour of non-use in calibration mode; with last reading reminder |
| Environment | 0 to 50°C (32 to 122°F); RH max 95% non-condensing |
| Weight | 360g (12.7oz.) |
| Dimensions | 193 x 104 x 69 mm (7.6 x 4.1 x 2.7") |
Click on a category to view a selection of compatible accessories with the Hanna HI96771 Ultra High Range Chlorine Portable Photometer.
| Range | Free Chlorine (P1): 0.00 to 5.00 mg/L (ppm); Ultra High Range Chlorine (P2): 0 to 500 mg/L (ppm) |
| Resolution | Free Chlorine (P1): 0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/L; Ultra High Range Chlorine (P2): 1 mg/L from 0 to 200 mg/L; 10 mg/L above 200 mg/L |
| Accuracy @ 25°C (77°F) | Free Chlorine (P1): ±0.03 mg/L ±3% of reading; Ultra High Range Chlorine (P2): ±3 mg/L ±3% of reading |
| Method | adaptation of the Standard Methods for the Examination of Water and Wastewater, 4500-Cl method |
| Auto-off | after ten minutes of non-use in measurement mode; after one hour of non-use in calibration mode; with last reading reminder |
| Environment | 0 to 50°C (32 to 122°F); RH max 95% non-condensing |
| Weight | 360g (12.7oz.) |
| Dimensions | 193 x 104 x 69 mm (7.6 x 4.1 x 2.7") |
Click on a category to view a selection of compatible accessories with the Hanna HI96771 Ultra High Range Chlorine Portable Photometer.