Features a revolutionary data maintenance and storage concept for automated test sequences and individual measurements up to 50,000 data records. Housed in a compact, impact resistant casing with integrated rubber protector.
Features a revolutionary data maintenance and storage concept for automated test sequences and individual measurements up to 50,000 data records. Housed in a compact, impact resistant casing with integrated rubber protector.
Discontinued!
This product has been discontinued and is no longer available. There is no direct replacement. Don't worry, we have some options to help you find what you need:
Visit our Voltage Testers category for similar products.
Or let's do a search and help you find what you need:
Test Instrument for Testing Electrical Devices, Medical Devices and Welding Units - with Test Sequences per VDE 0701-0702, IEC 62353 and IEC 60974-4
Features
Sequence Designer
User-defined test sequences can be created at the PC and transferred to the test instrument.
Report Designer
Report templates can be created for the printer Z721S at the PC and transferred to the test instrument.
Display with Selectable Language
The display panel consists of a backlit, color multi-display at which menus, setting options, measurement results, instructions and error messages, as well schematic and wiring diagrams appear.
Data Entry
Data can be entered, for example, via a barcode reader connected to the USB port, a RFID scanner, a USB keyboard, or via the softkey keyboard when it appears at the display. The touch screen of SECUTEST PRO (or devices with Feature E01) allows for the convenient entry of data and comments while menu control is still based on softkeys.
Creating a Database
A complete test structure with data regarding customers, buildings, floors, rooms and test objects can be created in the test instrument. This structure makes it possible to assign single measurements or test sequences to devices under test belonging to various customers. Manual single measurements can be grouped together into a so-called "manual sequence".
The SECUTEST PRO test instruments and those instruments with database expansion (Feature KB01) enable the user to prepare a test structure by means of the ETC (Electric Testing Center) software at the PC for subsequent transmission to the test instrument.
Software Update
The test instrument can always be kept current thanks to firmware which can be updated via the USB slave port. Software is updated during the course of recalibration by our service department, or directly by the customer.
Report Generating Functions
All of the values required for approval reports or device logbooks for electrical equipment (e.g. per ZVEH) can be measured with this instrument. The measured data can be documented and archived thanks to the measurement and test report that can be printed with a thermal printer connected to the USB port, or stored to a PC.
Automatic Detection of Measuring Point Changes
During protective conductor measurement, the test instrument recognizes whether or not the test probe is in contact with the protective conductor, which is indicated by means of two different acoustic signals. This function is very useful where several protective conductor connections need to be tested.
Mains Connection Analysis
Line voltage and frequency are measured and compared with the data specified in the setup menu. Momentary voltage or nominal voltage in accordance with the standard is required, for instance in order to extrapolate measured values for the leakage current measurement.
Automatic Detection of Mains Connection Errors
The device automatically recognizes mains connection errors if the conditions in the following table have been fulfilled. The user is informed of the type of error, and all measuring functions are disabled in the event of danger.
Applications
Ideal for the following applications: Appliance Testing, Electrical Medical Devices, Arc Welding Equipment.
| Protection Class | I per IEC 61010-1/EN 61010-1/VDE 0411-1 |
| Nominal Voltage | 230V |
| Test Voltage | 2.3kV AC 50Hz or 3.3kV DC (mains circuit/test socket to mains PE terminal, USB, finger contact, probe, test socket) |
| Measuring Category | 250V CAT II |
| Pollution Degreee | 2 |
| Line Voltage | 230V AC ±0.2% |
| Line Frequency Waveform | 50Hz ±2Hz |
| Ambient Temperature | +23°C ±2K |
| Protection Class | I per IEC 61010-1/EN 61010-1/VDE 0411-1 |
| Nominal Voltage | 230V |
| Test Voltage | 2.3kV AC 50Hz or 3.3kV DC (mains circuit/test socket to mains PE terminal, USB, finger contact, probe, test socket) |
| Measuring Category | 250V CAT II |
| Pollution Degreee | 2 |
| Line Voltage | 230V AC ±0.2% |
| Line Frequency Waveform | 50Hz ±2Hz |
| Ambient Temperature | +23°C ±2K |