Featuring automatic or manual setups and preprogrammed monitoring modes, with 4 differential 50/60hz, AC/DC voltage channels and 4 AC/DC current channels with flex probe power supply.
Our rentals come fully calibrated to traceable standards. Have your own instrument? We also offer calibration services for this model (CERTIWM3).
Featuring automatic or manual setups and preprogrammed monitoring modes, with 4 differential 50/60hz, AC/DC voltage channels and 4 AC/DC current channels with flex probe power supply.
Our rentals come fully calibrated to traceable standards. Have your own instrument? We also offer calibration services for this model (CERTIWM3).
All-in-one power and energy/demand monitoring tool that provides you with everything needed to get the job done. In addition, this advanced power quality analyzer has more user-centric power quality management functionalities and versatile communication options.
Features
Applications
Measured Parameters | |
Differential Voltage (4): 512 samples/cycle, 16 bit resolution | 1 to 1000 Vrms, AC/DC; ±0.1% reading, <40 V ±0.05% FS IEC 61000-4-30 Class A: 60-1000Vrms, ±0.1% of Udin, range of 10 to 150% of Udin Transients: 0 ro 1414 Vpk, ±0.2% of Udin Transients - High speed: 1 MHz (Xplorer/400), 10 to 2000 Vpk, ±10% of reading, ±0.5% FS |
Current Inputs with CTs (4): 512 samples/cycle, 16 bit resolution | Range probe dep., AC/DC, ±0.1% reading ±0.05% FS |
Frequency Range | 16 to 25 Hz, 42.5 to 69 Hz, ±0.01 Hz |
Calculated Parameters | |
Distortion - 200 MS, 3 Seconds, 10 Minute Windows | Vthd: 0 to 100%, ±5% for V ≥ 1% Vnom V Ind Harm: DC, 2 to 127, ±5% for V ≥ 1% Vnom Ithd: 0 to 100%, ±5% for V ≥1% Vnom I Ind Harm: DC, 2 to 63, ±5% for V ≥1% Vnom |
Power/Energy - 1 Second Sampling | Real Power (W) - P: meets 0.2 second requirements, range probe dep. Apparent Power (VA) - S: meets 0.2 second requirements, range probe dep. Reactive Power (var) - Q: meets 0.2 second requirements, range probe dep. Power Factor (W/VA) -"true" -1 to 0 to +1 Displacement PF -1 to 0 to +1 Demand (in W): meets 0.2 second requirements, range probe dep. Energy (in Wh): meets 0.2 second requirements, range probe dep. |
Miscellaneous | Pst: 10 minutes: 0.2 to 10, ±0.05 at Pst=1 Plt: 2 hours: 0.2 to 10, ±0.05 at Pst=1 |
Standards Compliance | |
Power Quality | IEC 61000-4-30 Class A Edition 3 Compliant IEEE 1159: 2009 |
Power | IEEE 1459: 2000 |
Harmonics | IEC 61000-4-7 Class 1: Edition 2 (2008) IEEE 519: 2014 |
Voltage Flicker | IEC 61000-4-15: Edition 2 (2010) IEEE 1453: 2011 |
Compliance/Testing | EN 50160: 2010 |
General Specifications | |
Display | 7" (17.78 cm) WVGA color graphic, icon-based touch LCD, LED backlit |
Environmental Conditions | Operating temperature: 0 to 50°C (32 to 122°F) Storage temperature: -20 to 55°C (-4 to 131°F) Humidity: 10 to 90% non condensing |
Power | Supply: AC Adapter, 90 to 264 V AC 50/60 Hz Run time on full charge: 3 hours (3 hours charge time) |
Dimensions | 10 x 2.75 x 8" (25.4 x 7.0 x 20.3 cm) |
Weight | 4.2 lbs (1.9 kg) |
Measured Parameters | |
Differential Voltage (4): 512 samples/cycle, 16 bit resolution | 1 to 1000 Vrms, AC/DC; ±0.1% reading, <40 V ±0.05% FS IEC 61000-4-30 Class A: 60-1000Vrms, ±0.1% of Udin, range of 10 to 150% of Udin Transients: 0 ro 1414 Vpk, ±0.2% of Udin Transients - High speed: 1 MHz (Xplorer/400), 10 to 2000 Vpk, ±10% of reading, ±0.5% FS |
Current Inputs with CTs (4): 512 samples/cycle, 16 bit resolution | Range probe dep., AC/DC, ±0.1% reading ±0.05% FS |
Frequency Range | 16 to 25 Hz, 42.5 to 69 Hz, ±0.01 Hz |
Calculated Parameters | |
Distortion - 200 MS, 3 Seconds, 10 Minute Windows | Vthd: 0 to 100%, ±5% for V ≥ 1% Vnom V Ind Harm: DC, 2 to 127, ±5% for V ≥ 1% Vnom Ithd: 0 to 100%, ±5% for V ≥1% Vnom I Ind Harm: DC, 2 to 63, ±5% for V ≥1% Vnom |
Power/Energy - 1 Second Sampling | Real Power (W) - P: meets 0.2 second requirements, range probe dep. Apparent Power (VA) - S: meets 0.2 second requirements, range probe dep. Reactive Power (var) - Q: meets 0.2 second requirements, range probe dep. Power Factor (W/VA) -"true" -1 to 0 to +1 Displacement PF -1 to 0 to +1 Demand (in W): meets 0.2 second requirements, range probe dep. Energy (in Wh): meets 0.2 second requirements, range probe dep. |
Miscellaneous | Pst: 10 minutes: 0.2 to 10, ±0.05 at Pst=1 Plt: 2 hours: 0.2 to 10, ±0.05 at Pst=1 |
Standards Compliance | |
Power Quality | IEC 61000-4-30 Class A Edition 3 Compliant IEEE 1159: 2009 |
Power | IEEE 1459: 2000 |
Harmonics | IEC 61000-4-7 Class 1: Edition 2 (2008) IEEE 519: 2014 |
Voltage Flicker | IEC 61000-4-15: Edition 2 (2010) IEEE 1453: 2011 |
Compliance/Testing | EN 50160: 2010 |
General Specifications | |
Display | 7" (17.78 cm) WVGA color graphic, icon-based touch LCD, LED backlit |
Environmental Conditions | Operating temperature: 0 to 50°C (32 to 122°F) Storage temperature: -20 to 55°C (-4 to 131°F) Humidity: 10 to 90% non condensing |
Power | Supply: AC Adapter, 90 to 264 V AC 50/60 Hz Run time on full charge: 3 hours (3 hours charge time) |
Dimensions | 10 x 2.75 x 8" (25.4 x 7.0 x 20.3 cm) |
Weight | 4.2 lbs (1.9 kg) |