This product is a load cell that has full-bridge strain gauge sensors complete with 15-pin plug. This load cell is used with the Chatillon DFS-R-ND Series digital indicator and they may also be used with the Chatillon TCD Series digital force testers (recalibration with a TCD Series Console is required).
This product is a load cell that has full-bridge strain gauge sensors complete with 15-pin plug. This load cell is used with the Chatillon DFS-R-ND Series digital indicator and they may also be used with the Chatillon TCD Series digital force testers (recalibration with a TCD Series Console is required).
Chatillon SLC-1000 has an accuracy up to 0.1% FS.
Features
| Lbf | 1000 x 0.01 |
| Kgf | 500 x 0.05 |
| N | 5000 x 0.2 |
| Deflect | 0.012 +0.003" |
| Accuracy | +0.25% FS |
| Input Resistance | 1000 ohms ±100 |
| Output Resistance | 1000 ohms ±100 |
| Rated Output | 2.0 mV/V |
| Safe Overload | 150% Full Scale |
| Non-Linearity | ±0.05% Full Scale |
| Hysteresis | ±0.05% of Full Scale |
| Zero Balance | ±0.1% of Full Scale |
| Operating Temperature | 40 to 110°F (5 to 45°C) |
| Instrument Weight | 1 lb (0.5 kg) |
| Shipping Weight | 1.5 lbs (0.7 kg) |
| Lbf | 1000 x 0.01 |
| Kgf | 500 x 0.05 |
| N | 5000 x 0.2 |
| Deflect | 0.012 +0.003" |
| Accuracy | +0.25% FS |
| Input Resistance | 1000 ohms ±100 |
| Output Resistance | 1000 ohms ±100 |
| Rated Output | 2.0 mV/V |
| Safe Overload | 150% Full Scale |
| Non-Linearity | ±0.05% Full Scale |
| Hysteresis | ±0.05% of Full Scale |
| Zero Balance | ±0.1% of Full Scale |
| Operating Temperature | 40 to 110°F (5 to 45°C) |
| Instrument Weight | 1 lb (0.5 kg) |
| Shipping Weight | 1.5 lbs (0.7 kg) |